![Satec PM130E Installation And Operation Manual Download Page 28](http://html.mh-extra.com/html/satec/pm130e/pm130e_installation-and-operation-manual_1200370028.webp)
Chapter 4 Setup Menus
21
4.2 Communications Port Setup Menu
This menu allows you to access the communications port options
that the
PM130
uses to communicate with a master computer.
Table 4-2 lists the communications options, their code names and
applicable choices.
Activate the middle window to scroll through the list of available
options, and then activate the lower window to set the option
value.
To select and view a setup option:
Press to activate the middle window.
Use the up/down arrow keys to scroll to the desired option.
The option setting will appear in the lower window.
To change the selected option:
Press to activate the lower window.
Press the up/down arrow keys to scroll to the desired
value.
Press to store the selected value or press
to quit the setup menu.
Table 4-2 Communications Options
(
∗
default setting)
Code Parameter
Options
Description
Prot
Communications
protocol
ASCII
∗
rtu
dnP3
ASCII protocol
Modbus RTU protocol
DNP3.0 protocol
rS
Interface standard
485
RS-485 (not changeable)
0
∗
-99 ASCII
Powermeter address
Addr
Address
1
∗
-247 Modbus
0
∗
-255 DNP3.0
bAud
Baud rate
110
110 baud
300
300 baud
600
600 baud
1200
1200 baud
2400
2400 baud
4800
4800 baud
9600
∗
9600 baud
19.20
19,200 baud
dAtA
Data format
7E
7 bits, even parity
8n
∗
8 bits, no parity
8E
8 bits, even parity
SELECT
Port
ASCII
Prot
SELECT
CHG
ENTER
Port
ENTER
SELECT
ENTER
ESC
Summary of Contents for PM130E
Page 1: ...Se Series PM130 Powermeters Installation and Operation Manual BG0306 Rev A1 ...
Page 2: ...Series PM130 Powermeters Installation and Operation Manual ...
Page 6: ...iv Quick Start TYPICAL INSTALLATION see Section 2 2 4 for full instructions Wiring Mode 4LL3 ...
Page 12: ...Chapter 1 Introduction 5 Instrument Dimensions Figure 1 1 PM130 Dimensions ...
Page 14: ...Chapter 2 Installation 7 Figure 2 2 The PM130 Terminations ...
Page 17: ...Chapter 2 Installation 10 Figure 2 4 Current Input Connection ...