12
Operating Instructions YTM15MA
Operation
Refer also to the following flow diagram for the
procedure .
Device should be
tested
Run
performance Test
Test OK?
Device is OK .
Heating Module
Calibration/
Adjustment
Weighing system
tested?
Test failed
2x?
Weighing System
Calibration/
Adjustment
no
no
no
yes
yes
yes
Fig . 3-2: Process of testing the heating module MA37 | MA160