Operating Instructions YTM15MA
17
Operation
t
Remove the measurement cable connector from
the temperature measuring device .
t
Allow the device to cool down .
t
Open the hood .
t
CAUTION
Danger of injury due to hot surfaces!
Observe the information in Chapter 4 .2, Page 9 .
Remove the adjustment disk from the pan retainer
using the disk holder (removal tool) and leave to
cool on a temperature-resistant pad .
y
The calibration of the heating module is now
complete .
4.5 Calibrate/Adjust Heating Module
(MA37 | MA160)
Calibration and adjustment of the heating module
involves calibration on two fixed temperature settings .
The heating module can be adjusted directly
afterwards depending on the result of the calibration .
The result is saved in the calibration report .
Carry out the calibration at the subsequent installation
site in order to ensure precise calibration of the
heating module . The calibration takes around
100 minutes .
System Requirements
− At least 30 minutes have passed since the last
application measurement .
− The device has cooled down with the hood open .
Procedure
t
Call up the setup menu [Settings] .