Operating Instructions YTM15MA
19
Operation
t
Place the connector (1) of the
measurement cable into the
temperature measuring device .
t
NOTICE
Observe the information in
Chapter 4 .2, Page 9 . Use the disk
holder to position the adjustment
disk (2) on the pan retainer .
t
Close the hood .
t
Press the [START] key .
y
The heating module will heat up .
The device requires a heating-up time of
45 minutes .
t
Wait for the end of the heating-up
phase .
y
The time elapsed is shown at the top
of the display .
The heating-up phase is followed by
a stabilization phase (specification in
percent) .
t
Wait for the end of the stabilization
phase .
y
After the end of the stabilization
phase, the device emits a beep .
1
2
Heating Module
Adjustment
Remove pan,
position adjustment disk and close hood.
1 Target temperature is 100°C
Press START.
w
02:08
Heating Module
Adjustment
Heating temperature 100°C
Heating-up phase: 45:00 min
42:56
Heating Module
Adjustment
Heating temperature 100°C
Stabilization phase: 23%