Operating Instructions YTM15MA
13
Operation
Reasons for an incorrect measurement may be:
− Non-compliance regarding ambient conditions .
− Intense direct sunlight .
− Strong draft .
− "ReproEasy Pad" was not used correctly during the
performance test . E .g . the protective film was not
removed from the "ReproEasy pad" or the incorrect
side of the "ReproEasy pad" was placed on the pan .
− Weighing system is defective .
4.4 Calibrate Heating Module
(MA37 | MA160)
Calibration of the heating module involves calibration
at a fixed temperature setting . The result is saved in
the calibration report .
Carry out the calibration at the subsequent installation
site in order to ensure precise calibration of the
heat ing module . The calibration takes at least
50 minutes .
System Requirements
− At least 30 minutes have passed since the last
application measurement .
− The device has cooled down with the hood open .