16
Operating Instructions YTM15MA
Operation
t
Switch on the temperature
measuring instrument .
After the heating-up and stabilization
phase, a total of six temperature values
must be read off and noted down .
t
Read and note down the first
temperature value .
After a minute, there is another beep
signaling that it is time to read off the
next temperature value .
t
Calculate the average temperature
(mean value) .
t
Press the [ . . .] key .
t
Enter the calculated average
temperature in the following input
screen .
y
The result of the calibration is
displayed .
The result of the calibration can
subsequently be called up at any time
via the menu Settings ->
Calibration/Adjustment -> Calibration
Report -> Heating Calibration .
t
Switch off the temperature
measuring instrument .
45:12
Calibrating the heating
Heating temperature 140°C
Please read the temperature from the
display of the heating adjustment set.
Please read off the temperature: 1
Average temperature
Calculate the mean value
(T1+T2+T3+T4+T5+T6)/6 and press ...
for the input.
Calibrating the heating
Date and time
2016-04-06 08:21
Target value
140°C
Actual value
143.20°C
Deviation
3.20°C