8
Operating Instructions YTM15MA
Operation
Pos.
Product
Quantity
6
Measurement cable (permanently attached to the
adjustment disk)
1
Not illustrated
Instructions for the Heating Adjustment Set
Not illustrated
Instructions for the Temperature Measuring Instrument
Not illustrated
Factory calibration certificate with serial number
(ref . ID no .)
3.2 Inserting the Battery
t
Insert the battery into the temperature measuring instrument (to do this, refer to
the Instructions for the Temperature Measuring Instrument) .
4
Operation
4.1 Notes on Calibrating the Moisture Analyzer
When calibrating the moisture analyzer using the heating adjustment set, there is an
examination of whether the moisture analyzer is adjusted to the ambient conditions
at the place of use . A basic heating adjustment has already been carried out at the
factory . This adjustment is valid for many installation sites under laboratory
conditions .
However, if you have particular ambient conditions, e .g . strong direct sunlight or
operation in an exhaust, then an inspection will be necessary .
The adjustment disk has a special coating, which absorbs almost all heat radiation .
This ensures that your measurements are reliably adjustable to all operating sites .
If calibration is carried out, then this must be calibrated in the ambient conditions
present . E .g .: If the moisture analyzer is operated in a room with exhaust, then the
moisture analyzer must also be calibrated at this place of use .