Operating Instructions YTM15MA
9
Operation
The following conditions should be taken into account
when calibrating the moisture analyzer:
− Carry out calibration in a temperate area . The room
temperature should be 20°C .
− Carry out calibration in a clean measurement
location .
− Use a stable and level surface (table) .
− Drafts, e .g . from open doors/windows, should be
avoided .
If required: Shield the device .
− The frequent movement of persons in front of the
device should be avoided (unstable conditions) .
− Avoid direct sunlight .
− Avoid gases or dusts .
− Establish a safety distance in the direct vicinity of
the device . At least 20 cm on all sides and at least
1 m above the device .
4.2 Transporting and Removing the
Adjustment Disk
CAUTION
Danger of burns due to hot adjustment disk!
t
Always use the disk holder (removal tool) to
transport the adjustment disk or remove it from
the moisture analyzer after calibration .
t
Allow the adjustment disk to cool on a clean,
temperature-resistant pad before or after
completion of a measurement .
t
Only place the adjustment disk back into the
carrying case after cooling .
t
Wear personal protective equipment .