26
Calibration with a User-defined Weight
You can define any weight as the calibration weight.
The weight value of this weight is stored automatically in the
INPUT
mode
under the “
cal./adj. wt.
” menu item.
– Activate the service mode (see page 11).
– Perform external or internal span adjustment.
– Press F6
F6
F6
F6
F6 to select the “
det. user wt.
” menu item.
– Tare the balance if necessary, and then press F6
F6
F6
F6
F6 to start the calibration
routine.
– The last calibration weight that was entered in the INPUT mode is dis-
played.
– Center the calibration weight you wish to define on the weighing pan.
– The deviation from the stored weight is displayed.
– The new user weight can be selected and stored by pressing
F6.
– Press
F1
to cancel the routine.
– The balance returns to the SELECT mode.
reproTEST
This tests the reproducibility of the automatic application of the built-in calibra-
tion weight.
The internal weight is applied, measured and removed 6 times.
At the end of the routine, the reproducibility value is displayed.
High Resolution - Display * 10
The balance switches to SuperRange resolution and an additional digit is dis-
played (for test purposes and initial verification).
La-037e.BMP
La-036e.BMP
La-039e.BMP
La-043e.BMP
La-042e.BMP
Summary of Contents for Master Pro LA12000P
Page 50: ...50 Spare part list LP LA LP LA Exploded View Diagram 1 LP_exp_1 WMF LA LP...
Page 52: ...52 LP LA Exploded View Diagram 2 LP_exp_2 EPS...
Page 54: ...54 LP LA Exploded View Diagram 3 LP_exp_3 EPS...
Page 56: ...56 LP_exp_4 EPS LP LA Exploded View Diagram 4...
Page 58: ...58 LP_exp1a WMF LP LA Exploded View Diagram 1a LP LA...
Page 62: ...62 LP_exp_3a WMF LP LA Exploded View Diagram 3a 301 302 308 309 305 306 307 304 305...