OPTICAL TUNABLE FILTER OTF-920
11-1
11
Troubleshooting
If after checking the above, the unit is still not operating correctly please contact Santec's technical
support division.
Problem
Possible Cause
Remedy
Power cable is unplugged.
Make sure the cable is plugged in.
Fuse blown.
No fuse.
Wrong supply voltage.
ENT has not been pressed.
Press ENT key after entering values.
Unit is in power setting mode.
Press the Wavelength / Power
selector key to change to
wavelength setting mode.
Setting is out of the tuning range.
Check the specification of unit.
Unit is in calibration mode.
Exit calibration mode.
Attenuator option not fitted.
Check the specification of unit.
Unit is in wavelength setting mode.
Press the Wavelength / Power
selector key to change to power
setting mode.
Setting is out of the tuning range.
Check the specification.
Unit is in calibration mode.
Exit calibration mode.
ENT has not been pressed.
Press ENT key after entering values.
Unit is in APC mode.
Press APC key to cancel APC mode.
Unable to use front
panel.
In GPIB remote mode.
If the GPIB remote LED is lit, press
local key to release the GPIB remote
mode. Note this function will not work
if the Front Panel lockout command
has been issued via GPIB.
Wavelength of input light is different
from wavelength setting of filter.
Set wavelength or use peak search
to scan for wavelength peak.
Attenuation is set too high.
Reduce the attenuation setting or
set the unit to APC mode with
appropriate setting.
GPIB address is not set correctly.
Press Shift +Local to check GPIB
address and set if necessary.
Delimiter setting is incorrect.
Check the delimiter setting.
Cable is not connected properly.
Check the connection of the cable.
Communication protocol is incorrect. Check the communication protocol.
Delimiter setting is incorrect.
Check the delimiter setting.
Cable is not connected properly.
Check the connection of the cable.
Cable is not wired correctly.
Check the pin assignment of the
cable.
RS-232C Communication
is not possible.
Open the fuse holder cover on the
rear panel and check the supply
voltage and fuse.
Unable to switch on.
Unable to change
wavelength setting.
GPIB Communication is
not possible.
Unable to change
attenuation setting.
No light output.