8-2
OTF-920 Operation Manual
8.2. RS-232C Control
The delimiter for RS-232C communication is set in the factory as CR (carriage return). This can be
changed when power to the OTF-920 is initially switched on. For an explanation of this procedure
please see Section 7.1 Power On.
The following table lists the communication conditions that should be set to enable communication
through the RS-232C interface of the OTF-920.
The OTF-920 provides support for RS-232C communication through the SIF connector on the front
panel. The RTDX (transmit) terminal of the OTF-920 should be connected to the receive terminal of
the external device (e.g. computer). The RRDX (receive) terminal of the OTF-920 should be
connected to the transmit terminal of the external device. Please ensure that the handshake line on the
external device is closed. The terminals for the SIF interface (Hirose Electric HR10-7P-6S) are listed
below:
1
N / C
Not used (Do not connect)
2
N / C
Not used (Do not connect)
3
G N D
Ground
4
N / C
Not used (Do not connect)
5
R T D X
RS-232C Transmit
6
RRDX
RS-232C Receive
Communication Method
Full Duplex
Baud rate
9600 bps
Data Length
8 bit
Stop Bit
1 bit
Parity
N o n e
Flow Control
N o n e
1
2
3
6
5
4
Viewed from front of OTF-920