Operating Manual Q-Jet CT20
© Sanosil
5/32
2.2 Improper use
Always and only use Sanosil disinfectants with a maximum hydrogen peroxide content of 7.5%. The use of any
other disinfectants or accessories that have not been expressly approved by Sanosil Ltd. violates the rules de-
scribed in this manual and may result in serious injury as well as damage to the device. The device is not in-
tended for use as the only disinfection procedure. Instead, it is intended as a supplement to regular cleaning
and disinfection.
DANGER
Risk of explosion! / Risk of injury!
No flammable liquids may be nebulised!
Only Sanosil disinfectants may be nebulised.
2.3 Dangers and warnings
The device must be used in full compliance with this operating and maintenance manual.
•
It is strictly forbidden to use the device in an explosive or flammable atmosphere, in dusty areas or in the
presence of people or animals. Prevent people from intentionally or accidentally entering the room during
the fogging process or during the waiting time afterwards.
•
There is a risk of injury if unauthorised accessories are used!
•
For this reason, only use original accessories from Sanosil Ltd.
•
Place the unit on a stable, level, dust-free, non-slip, horizontal surface with a load-bearing capacity suitable
for its weight.
•
Always disconnect the device from the power supply before removing the rear panel.
•
Carefully read all the information provided on the label of the Sanosil disinfectant.
2.4 Safety instructions for the aerosol and device:
Nebulised disinfectant/aerosol:
•
When using the Sanosil Q-Jet and Sanosil disinfectant, high concentrations of hydrogen peroxide are pro-
duced in the air. These can be detrimental or hazardous to health. Only operate the device in rooms that
have been prepared for disinfection and are sealed off. Prevent people or animals from entering rooms
where active aerosol disinfection is present. Prevent aerosols containing peroxide from escaping into parts
of the room with unprotected persons during and for some time after disinfection.
•
Once the fogging process using the Sanosil Q-Jet has been completed, the air will be full of fogged Sanosil
disinfectant. DO NOT ENTER the treated room FOR AT LEAST 120 MINUTES AFTER TREATMENT. The wait-
ing time of 120 minutes must be observed. This does not mean, however, that the room can be safely re-
entered after this time. The maximum limit for hydrogen peroxide in the air without any protective equip-
ment is 1 ppm.
•
If someone needs to enter the treated room before the end of the waiting time or if the hydrogen perox-
ide content must be measured, the individuals concerned must use personal protective equipment. This
includes a full-face, sealed respirator (filter class ABEK P3) and gloves if necessary.
•
The device is intended for use with Sanosil disinfectant (S010 / S015). Do not use any other disinfection
products or liquids.
•
ONLY use Sanosil disinfectant to ensure maximum safety and effectiveness. The use of any other cleaning
agent or disinfectant may cause serious injury and environmental damage.
•
No one may be permitted to enter rooms while the equipment is in use. Warning signs must be placed at
the entrances.