Operating Manual Q-Jet CT20
© Sanosil
11/32
•
"Countdown timer set" allows you to set the
countdown timer, i.e. how much time elapses after
the program is started and the fogging
process is started
•
The "Minimum contact time" controls how long
the device continues to issue a warning message
after the spraying process has ended
4.2 Disinfection mode
The CT 20 features two different operating modes: "Auto" and "Expert". The difference is how the disinfectant is
delivered into the room.
4.2.1 Auto mode
Auto mode is designed to deliver any amount of disinfectant into the room such that condensation is avoided
as much as possible. Condensation means that the air is saturated with moisture and the liquid starts to settle
on the coldest surfaces. This allows drops to form there. Condensation has an adverse effect on the disinfection
result, since it prevents the disinfectant from being distributed as evenly as possible in the room and from cov-
ering all surfaces evenly as a relatively dry mist.
In addition, condensation on sensitive surfaces can result in unwanted effects. Auto mode prevents condensa-
tion and activates a humidity sensor that constantly measures the relative humidity. As soon as there is a risk of
condensation forming, Auto mode stops the aerosol input and waits until the air humidity has dropped again to
such an extent that the air can absorb more liquid.
The program then continues until the entire desired dose has been supplied to the room without condensation.
Auto mode is ideal for straightforward and uncomplicated maintenance disinfection, e.g. in hotel rooms, gyms,
restaurants and laboratories, etc.
•
Press "Run" to treat a new room or to load a saved
room profile.
•
Press "Quick Action" to treat a new room