Operating Manual Q-Jet CT20
© Sanosil
24/32
6.5 After disinfection
•
Wait for the necessary disinfection time as well as the
minimum decay time for the disinfectant to break
down. (See section 5: usage profile)
•
Do not enter the room during the decay time
•
If you selected usage profile C: after the end of the
spray cycle, switch on the supply air and extract air
ventilation at the highest level after the minimum de-
cay time of 120 minutes in order to expel any resid-
ual H
2
O
2
ATTENTION
Health risk
If someone must re-enter the room before the treatment process is completed or if an H
2
O
2
meas-
urement must be taken, personal equipment (PES), including a full-face respirator (filter class ABEK
P3) and gloves must be worn.
•
As soon as the level of nebulised hydrogen perox-
ide in the air has decreased below 1 ppm, the
room can be used again without any restrictions