33
-
12.Miscellaneous
12
‑
1
How to use the Moisture reading checker(option)
Press the electrode against the checker as shown in the following sketch in power ON condition.
When the reading of the Meter matches the figure on the checker, the Meter's reading is acceptable.
When the reading deviates from the checker's, please request our agent or the nearest our branch to
calibrate the Meter.
※
An object for checking by the Moisture reading checker differs face side from back side.
Use after surely confirming model, type of the Meter and measurement mode.
※
When checking the Meter, carry out after turning the temperature compensating function OFF.
Refer to 7.TEMPERATURE COMPENSATION for the details.
12
‑
2
Cover for the needle electrode (Fitting/removing methods of the cover)
Fit the cover in the needle electrode spreading out by fingers holding both sides of the cover for
the electrode.
When the cover is tight, fit by force prying it.
Carry out the same procedures for removing the cover.
Fitting / removing the cover by force causes an injury occasionally.
Be careful not to be injured with the needle electrode.
Ver.5 November 2007