Guidance and Manufacturer’s Declaration – Immunity All
ME Equipment and ME Systems
The
sam
is intended for use in the electromagnetic environment
specifi ed below. The customer or user of the
sam
should ensure
that it is used in such an environment.
Immunity Test
IEC 60601
Compliance
Electromagnetic
Test Level
Level
Environment - Guidance
ESD
+6kV Contact
+6kV Contact
Floors should be wood, concrete
IEC 61000-4-2
+8kV Air
+8kV Air
or ceramic tile. If fl oors are
synthetic, the r/h should be at
least 30%.
EFT 2kV
Mains
+2kV Mains
Mains power quality should be
IEC 610000-4-4
+1kV I/Os
+1kV I/Os
that of a typical commercial
or hospital environment.
Surge
+1kV Diff erential +1kV Diff erential Mains power quality should be
IEC 61000-4-5
+2kV Common
Common N/A
that of a typical commercial or
(no ground)
hospital environment.
Voltage
>95% Dip for
>95% Dip for
Mains power quality should be
Dips/Dropout
0.5 Cycle
0.5 Cycle
that of a typical commercial or
IEC 61000-4-11
60% Dip for
60% Dip
hospital environment. If the user
5 Cycles
5 Cycles
of the
sam
requires continued
30% Dip for
30% Dip for
operation during power mains
25 Cycles
25 Cycles
interruptions, it is recommended
>95% Dip for
>95% Dip for
that the
sam
be powered from
5 Seconds
5 Seconds
an uninterruptible power supply
or
battery.
Power
3A/m
3A/m
Power frequency magnetic fi elds
Frequency
should be that of a typical
50/60Hz
commercial or hospital
Magnetic Field
environment.
IEC 61000-4-8