Guidance and Manufacturer’s Declaration – Immunity
The
sam
is intended for use in the electromagnetic environment
specifi ed below. The customer or user of the sam should ensure
that it is used in such an environment.
Immunity Test
IEC 60601
Compliance
Electromagnetic
Test Level
Level
Environment - Guidance
Portable and mobile communications
equipment should be separated from
the
sam
by no less than the distances
calculated/listed below:
Conducted RF
3 Vrms
(V1) = 3 Vrms
D = (3.5/V1)(Sqrt P)
IEC 61000-4-6
150 kHz to
150 kHz to 80 MHz
80
MHz
Radiated RF
3V/m
D = (3.5/Eq)(Sqrt P)
IEC 61000-4-3
80 MHz to
(E1) = 3 V/m
80 to 800 MHz
2.5
GHz
D=(7/E1)(Sqet
P)
800 MHz to 2.5 GHz
Where P is the max power in watts and D
is the recommended separation distance
in meters.
Field strength from fi xed transmitters,
as determined by an electromagnetic
site survey, should be less than the
compliance levels (V1 and E1).
Interference may occur in the vicinity of
equipment containing a transmitter.