21
ZetrOZ, Inc.
Product Support
203-349-2798
11.6. PREVENT UNINTENTIONAL SHUTOFF
Once satisfi ed with the treatment settings, the
settings may be locked to prevent unintentional
shutoff during treatment by sliding the
Lock Switch
down into the ‘locked’ position as indicated by
the symbols on the
sam
Power Controller (see
Figure 11A). In this state the settings cannot be
changed until the
Lock Switch
is moved back to the
‘unlocked’ position.
Note: The system is not required to be locked in
order to operate. Always be aware of how
to disable a locked device. Always notify the
patient if their device is locked.
11.7. TREATMENT DELIVERY
Treatment will be delivered for the duration set by
the user. The treatment timer LEDs will count down
in 30 minute increments as treatment progresses.
Upon completion, the device will automatically shut
off and the ‘treatment complete’ LED will illuminate
for approximately 8 hours before turning off . See
section 12 for how to manually power ‘OFF’
the device.
Warning: If the
sam
Lock Switch is in the locked position, the
sam
Power Button will
not turn off the device until the
sam
Lock Switch is placed into the
unlocked position.
Warning: ALWAYS instruct the patient to alert the practitioner in the event of discomfort
during treatment.
Warning: As an additional safety measure, ALWAYS instruct the patient how to turn off
the
sam
Device and remove the
sam
Applicator in the event of discomfort
or emergency.
Warning: ALWAYS keep the
sam
Power Controller within easy reach so that power may
be ceased at any time.