10.3.2. DUAL-APPLICATOR MODE
The Y-adapter (optional) may be used to power two Applicators simultaneously. As
shown in Figure 8, connect the Y-adapter directly to the Power Controller wire and then
connect each Applicator to a wire jack at the other end of the Y-adapter.
When using two
sam
Applicators and two
sam
Ultrasound Coupling Bandages
simultaneously, the
sam
Ultrasound Coupling Bandages should be positioned so that
their footprints do not overlap.
Caution: DO NOT overlap
sam
Ultrasound Coupling Bandages (Figure 6).
Caution: DO NOT use the Y-adapter when using only one applicator for treatment.
Caution: Do NOT attach more than one Y-adapter to the
sam
Device.
Note: If two applicators will be used, two
sam
Ultrasound Coupling Bandages (one for
each applicator) and the Y-adapter must be used for the treatment.