6.17
Sweep
Pop up the soft menus for setting of sweep, including [Sweep Time], [Sweep Point], [Sweep Single], [Sweep Count]
[Sweep time Auto Manual]
Menu description: to change the sweep time of signal/spectrum analyzer by number keys, stepping keys or knob. When it is
under Manual mode, it indicates that sweep time can be set manually. When it is under Auto mode, the sweep time will be
automatically coupled based on resolution bandwidth, span and video bandwidth
The sweep time is coupled with the sweep type, resolution bandwidth, video bandwidth, average function and span. Changing
those parameters will result in change of the sweep time or triggering of the limit setting of sweep time.
[Sweep Continuous] / [ Sweep Single]
Menu description: to set continuous sweep or single sweep. Continuous sweep is the default measurement control method;
when enable [Sweep Single], sweep will be done only one time.