17
5.8
Use Pre-amplifier to Capture Weak Signals
Following example uses a S1103 signal generator and generates a 500MHz, -80dBm continuous signal.
Example
1)
Connect the signal generator output port to S3531 RF test port.
2)
Press
【
Preset
】
key to return the instrument to default status
3)
Set test frequency.
Press
【
FREQ
】
key
Select [Center FREQ] and input 500MHz
Press
【
Span
】
key
Select [Span] and input 10MHz
4)
Enable the pre-amplifier
Press
【
AMP
】
key
Select [Ref level] and input -20dBm
Select [Pre Amp On Off], enable the pre-amplifier
The trace will be displayed as shown in the figure below.
5)
Read out the result
Figure 4-12: Pre-amplifier On Measurement
5.9
Harmonic Distortion Measurement
Following example uses a S1103 signal generator and generates a 300MHz, -10dBm continuous signal.