Tel: 886.2.2175 2930 Email: [email protected]
www.salukitec.com
19
in Figure below.
Fig 5-14: TOI Distortion Measurement Connection
2)
Press
【
Preset
】
key to return the instrument to default status
3)
Set test frequency.
Press
【
FREQ
】
key
Select [Ctr FREQ] and input 500MHz
Press
【
SPAN
】
key
Select [SPAN] and input 5MHz
Press
【
BW
】
key
Select [Res BW Auto Man] and input 1kHz
4)
Use
【
Peak
】
and
【
Marker
】
keys to do the analyze.
Press
【
Peak
】
key, S3531 will search the MAX value and mark it with a marker 1.
Press
【
Marker
】
key
Select [Delta], a Marker 2 is activated and it is a Delta marker, the reference marker is Marker 1.
Press
【
Peak
】
key. Select [Next Peak].
5)
Readout the value of Marker 2. It is the TOI distortion