To review the data files refer to the Evoke SCS System Clinical Data Viewer User Manual.
7.7.1
Notes
During a programming session the data may be annotated with a time-stamped note.
1.
Type a text note in the
Notes
field
(see Figure 6.2).
2.
Touch the
Enter
button to save the text note with a time-stamp in the clinical data log
file, e.g.
. A screenshot of the CPA programming screen is also saved
in a ’Screenshot’ folder for the relevant visit on the CI.
7.7.2
Patient posture notes
1.
Touch the arrow in the
Notes
field
to display a list
of pre-defined notes (right).
2.
Select the required note to save it with a time-stamp in the clinical
data log file, e.g.
. A screenshot of the CPA
programming screen is also saved in a ‘Screenshot’ folder for the
relevant visit on the CI.
7.7.3
Screenshots
1.
Touch the
Communication Quality Indicator
icon (see Section 8.7.4) to save a
screenshot of the CPA programming screen.
2.
The icon will briefly change its appearance to show a camera graphic to indicate that
the screenshot was saved.
3.
The screenshot is saved in a ‘Screenshot’ folder for the relevant visit on the CI.
7.8
Neurophysiological measurements
The CPA can be used with the eCLS/CLS to measure the neurophysiological characteristics of
the patient’s spinal cord as well as to test the vertical alignment of the two implanted leads.
1.
Touch the
CLS
icon
, then touch the
More…
field and select
Show
Neurophysiological Measurements
. The neurophysiological measurement menu will
appear adjacent to the navigation menu icon (Figure 7.12).