Evoke
®
SCS System Clinical Manual
CLIN-UMAN-005175
Revision: 9.00
Page 20 of 101
Figure 6.10: You can drag the leads up, down or switch sides.
6.7
Programming overview
1. Test stimulation from several electrodes (see Section 6.8 and 6.9) and ask the patient to
report where they feel paresthesia.
2. Determine the stimulation electrodes and parameters that give the best paresthesia
coverage of the patient’s dermatomal pain area(s).
3. Determine the settings that provide a good measurement of the neural response (ECAP)
to stimulation (see Section 6.10).
4. Determine the settings for optimum closed-loop stimulation (see Section 6.11 and 6.13).
5. Save the program to the eCLS/CLS memory (see Section 6.15) and proceed to set up a
maximum of four programs.
6.8
Choose the stimulation electrodes automatically
1. Stop stimulation
if stimulation is running.
2. Change the
Electrode Selection Mode
icon (see Figure 6.2) until the
Automatic Electrode
Selection
icon
is displayed.
3. In the electrode display window, touch the electrode that you wish to use as the
stimulation electrode
.
4. Touch the stimulation electrode again to cycle through electrode configurations.
Tripolar: 1 lead
Bipolar: 1 lead
Tripolar: 2 leads
Bipolar: 2 leads