6
Clinical Interface – General programming using the CPA
6.1
Connecting the CI to the CST and eCLS/CLS
To program and download data from the eCLS or CLS, communication must first be established.
The CST connects to the CI to communicate using a radio signal transmitted to the eCLS or CLS.
1.
Turn on the CI and login to the ‘Saluda’ account.
2.
Check that the time and date on the CI are correct (bottom right corner of CI screen).
Refer to the CI instructions (see Section 5.1 for link) to change the time and date.
3.
Connect the CST to the CI with the USB connection. See Figure 6.1.
Figure 6.1: Connecting the CST to the CI.
4.
Position the CST near the stimulator and double-touch the Evoke Clinical Programing
Application (CPA) icon
to start the CPA (see Figure 6.2 for the main screen layout).
Note:
The CST can link with the stimulator within a 1.0 m (3.3 ft) radius.
Note
: Electromagnetic interference can affect the quality of this wireless link; see
Section 4.1.7 - Electromagnetic interference (EMI).