Evoke
®
SCS System Clinical Manual
CLIN-UMAN-005175
Revision: 9.00
Page 59 of 101
8
Clinical Interface – Settings guide
8.1
Stimulation settings menu
Stimulation Set
selection
Stimulation set slots, labeled A, B, C or D.
Empty slots represent unused stimulation
sets.
Stim Set Parameters
Set Defaults
Set stimulation settings to default values
for the selected stimulation set
Pulse Width
20 – 1000 μs
Program Parameters
Stimulation Frequency
(Hz)
10 – 1500 Hz closed-loop disabled
10 – 250 Hz closed-loop enabled
Maximum stimulation
current
1 – 50 mA
Current Increment
0.05 mA – 4.0 mA (depending on the
electrode setup)
Virtual Ground
- ClearCAP feature
Enable (checked) / Disable (unchecked)
Virtual Ground stimulation (ClearCAP). Sets
the Reference electrode as an input for
Virtual Ground (ClearCAP)
(see
Section 7.2).
Note:
Stim Set Parameters
only apply to the selected stimulation set.
Program
Parameters
apply to all stimulation sets. The available range of each setting is shown.