![SALUDA MEDICAL Evoke SCS System Clinical Manual Download Page 16](http://html2.mh-extra.com/html/saluda-medical/evoke-scs-system/evoke-scs-system_clinical-manual_3627820016.webp)
Evoke
®
SCS System Clinical Manual
CLIN-UMAN-005175
Revision: 9.00
Page 16 of 101
not already synchronized).
Note:
The time error between the CI and stimulator is displayed in brackets.
Note
: If out of sync, a red ‘clock’ icon will be displayed in the CLS icon window
and the time error between CI and stimulator is displayed in brackets in the CLS menu.
6.3
Check stimulator battery voltage
The CLS or eCLS battery voltage are indicated on the CLS icon
with a white bar and
voltage (3.25 to 4.10 V; Figure 6.4). If the battery voltage is low (<25 % full) the bar will be red.
Refer to the Evoke SCS System User Manual for instructions on charging the CLS battery or
Section 11.4 for charging the eCLS battery.
Figure 6.4: CLS icon with stimulator battery voltage indicator.
6.4
Retrieve or set the Patient ID and Patient Zone
6.4.1
For an existing patient
1.
On connection to the stimulator the Patient ID and Patient Zone are retrieved and
displayed under the CLS icon (see Figure 6.5).
Figure 6.5: CLS Icon with Patient ID and Zone underneath.
2.
Check that the correct Patient ID and Patient Zone have been retrieved.
6.4.2
For a new patient
1.
For a new patient, touch the
Patient ID
box under the CLS icon (see Figure 6.5).
2.
Delete any existing text and enter the new Patient ID.
3.
Touch the
Done
button that appears above the Patient ID when you finish typing (see
Figure 6.6).
Stimulator
battery
indicator
Patient ID
field
Patient
Zone field