![Rosemount Oxymitter 5000 Instruction Manual Download Page 31](http://html1.mh-extra.com/html/rosemount/oxymitter-5000/oxymitter-5000_instruction-manual_1487464031.webp)
Instruction Manual
IM-106-350C, Rev 2.2
July 2008
1-17
Hazardous Area Oxymitter 5000
NOTES:
High Sulfur Service:
High sulfur cell can be selected for any probe; add a line item note to your purchase order requesing the high sulfur ZrO2 cell in place of the standard ZrO2 cell.
(1)
Recommended uses: High velocity particulates in flue stream, installation within 10ft (3,5 m) of soot blowers or heavy salt cake buildup. Applications:
Pulverized coal, recovery boilers, lime kiln.
(2)
Where possible, specify ANSI or DIN designation; otherwise, provide details of the existing mounting plate as follows:
(3)
Startup, calibration, and operation can be implemented using the standard membrane keypad. Remote access and additional functionality available via
Fieldbus Communications (DeltaV).
Cont’d
Code
Operator Interface
(3)
1
Membrane Keypad - Fieldbus, Blind Cover
2
Membrane Keypad - Fieldbus, Window Cover
3
Gas Florescent LOI, Fieldbus, English only, Window Cover
Code
Language
1
English
2
German
3
French
4
Spanish
5
Italian
Code
Termination Filtering
00
No Option - Specified as part of Electronic Housing
Code
Calibration Accessories
00
No Hardware
01
Cal/Ref Flowmeters and Reference Pressure Regulator
02
Autocalibration Systems - order by seperate part number (for safe areas only)
Code
Hazardous Area Approval
00
certification selected elsewhere
Code
Control Suite Functionality
00
Basic Control Suite
01
Deduct Basic Control Suite
Code
Electronics to Probe Cable
00
No Cable, Integral Electronics
10
20 ft (6 m) Cable - Remote Electronics
11
40 ft (12 m) Cable - Remote Electronics
12
60 ft (18 m) Cable - Remote Electronics
13
80 ft (24 m) Cable - Remote Electronics
14
100 ft (30 m) Cable - Remote Electronics
15
150 ft (45 m) Cable - Remote Electronics
16
200 ft (61 m) Cable - Remote Electronics
Cont’d
1
3
00
01
00
00
00
Example
Plate with studs
Bolt circle diameter, number, and arrangement of studs; stud thread; and stud height above mounting plate.
Plate without studs
Bolt circle diameter, number, and arrangement of holes; thread; and depth of stud mounting plate with accessories.
Summary of Contents for Oxymitter 5000
Page 2: ......
Page 6: ......
Page 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...