Section 9
Maintenance and Service
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 9-1
Calibration with Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . page 9-1
FOUNDATION Fieldbus O
CAL Method . . . . . . . . . . . . . . page 9-6
Calibration with LOI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 9-6
Hazardous Area Oxymitter 5000 Repair . . . . . . . . . . . . . . page 9-8
OVERVIEW
This section identifies the calibration methods available and provides the
procedures to maintain and service the Hazardous Area Oxymitter 5000.
CALIBRATION WITH
KEYPAD
During a calibration, two calibration gases with known O
2
concentrations are
applied to the Hazardous Area Oxymitter 5000. Slope and constant values
calculated from the two calibration gases determine if the Hazardous Area
Oxymitter 5000 is correctly measuring the net concentration of O
2
in the
industrial process. A calibration record sheet is provided on the previous
page. Use photocopies of the calibration record sheet to track transmitter
performance.
Before calibrating the Hazardous Area Oxymitter 5000, verify that the
calibration gas parameters are correct by setting the gas concentrations
("Keys" in Section 5: Startup and Operation of Hazardous Area Oxymitter
5000 with Membrane Keypad or "LOI Menu Tree" in Section 6: Startup and
Operation of Hazardous Area Oxymitter 5000 with LOI) used when calibrating
the unit and by setting the calibration gas flowmeter.
The calibration gas flowmeter regulates the calibration gas flow and must be
set to 5 scfh. However, only adjust the flowmeter to 5 scfh after placing a new
diffuser on the end of the probe. Adjusting the flowmeter at any other time can
pressurize the cell and bias the calibration.
When working on this equipment on the laboratory bench, be aware that the Hazardous
Area Oxymitter 5000, probe tube, and flame arrestor hub can be hot [up to 300°C (572°F)]
in the region of the probe heater.
Install all protective equipment covers and safety ground leads after equipment repair or
service. Failure to install covers and ground leads could result in serious injury or death.
Summary of Contents for Oxymitter 5000
Page 2: ......
Page 6: ......
Page 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...