8. Slide the heater strut assembly into the probe per the instructions in
"Heater Strut Replacement", steps 9 through 14.
9. On systems equipped with an abrasive shield, install the dust seal
gaskets, with joints 180° apart.
10. Follow instructions in "Probe Installation" in Section 2: Installation to
install the Hazardous Area Oxymitter 5000 into the stack or duct. If there
is an abrasive shield in the stack, make sure the dust seal gaskets are in
place as they enter the 15° reducing cone.
Figure 9-11. Contact and
Thermocouple Assembly
Replacement
38740024
Strut
Tube
Supports
Contact And
Thermocouple
Assembly
Spring
Clip
Spring
Summary of Contents for Oxymitter 5000
Page 2: ......
Page 6: ......
Page 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...