Instruction Manual
IM-106-350C, Rev 2.2
July 2008
1-5
Hazardous Area Oxymitter 5000
Figure 1-2. Hazardous Area
Oxymitter 5000 Autocalibration
System Options
For systems with one or two Hazardous Area Oxymitter 5000 units per
combustion process, an optional SPS 4001B Single Probe Autocalibration
Sequencer can be used with each Hazardous Area Oxymitter 5000 to provide
automatic calibration gas sequencing. The sequencer performs
autocalibrations based on the CALIBRATION RECOMMENDED signal from
the Hazardous Area Oxymitter 5000, a timed interval set up in fieldbus, or
whenever a calibration request is initiated.
System Features
1. The CALIBRATION RECOMMENDED feature detects when the sensing
cell is likely out of limits. This may eliminate the need to calibrate on a
"time since last cal" basis.
2. The cell output voltage and sensitivity increase as the oxygen
concentration decreases.
HAZARDOUS AREA
OXYMITTER 5000
IMPS 4000
(1 to 4 Probes)
(Must be installed
in a safe area
or be X- or Z-purged
by the customer)
SPS 4001B
(1 Probe)
(Must be installed
in a safe area)
38740068
Summary of Contents for Oxymitter 5000
Page 2: ......
Page 6: ......
Page 14: ...Hazardous Area Oxymitter 5000 iv Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 82: ...Hazardous Area Oxymitter 5000 7 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 130: ...Hazardous Area Oxymitter 5000 9 24 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 140: ...Hazardous Area Oxymitter 5000 11 6 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 172: ...Hazardous Area Oxymitter 5000 B 2 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 192: ...Hazardous Area Oxymitter 5000 D 14 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 210: ...Hazardous Area Oxymitter 5000 E 18 Instruction Manual IM 106 350C Rev 2 2 July 2008 ...
Page 212: ...Instruction Manual IM 106 350C Rev 2 2 July 2008 Hazardous Area Oxymitter 5000 Index 2 ...