3
Installation
3.1
Installation requirements
The customer provides the following:
Installation Requirements
Table 3-1:
Customer provided
Specification
Mains supply (defined by customer)
110/120 Vac, 50 Hz
or
Mains supply (defined by customer)
220/240 Vac, 60 Hz
Circuit breaker
16A (maximum)
Mounting
Affix to floor or building structure
Compressed air
8-10 bar / 115 - 145 psi
For the electrical power wiring, use 16 AWG stranded, three conductor copper plain or tin
plated wire, rated for a minimum of 250 Vac for the required cable length. Cables must be
terminated in accordance with national/local electrical standards and must be rated for
operation in ambient temperatures > 176 °F (80 °C).
Emerson
™
does not supply a switching system with the instrument. The customer must
supply a suitably rated switch or circuit breaker to be included with this installation. Check
the installation of the switch for conformity in accordance with national/local regulations
and standards by inspection. The switch or circuit breaker must be suitably located, easily
reached, and identified as the disconnection device for the leak detection system.
The leak detection system provides a protective earth terminal. To prevent electrical shock
hazards, connect the instruments to a protective earth. Any interruption of the earth
connector inside or outside the instrument or disconnecting the earth terminal may cause
potential electrical shock hazard.
Installation
Reference Manual
9
Summary of Contents for CT4215
Page 1: ...Reference Manual 00809 0100 4245 Rev AB July 2018 Rosemount CT4215 Leak Detection System ...
Page 12: ...Contents ii Rosemount CT4215 ...
Page 28: ...Installation 16 Rosemount CT4215 ...
Page 30: ...Startup 18 Rosemount CT4215 ...
Page 44: ...Operation 32 Rosemount CT4215 ...
Page 50: ...Maintenance 38 Rosemount CT4215 ...
Page 54: ...Troubleshooting and diagnostics 42 Rosemount CT4215 ...