Program parameter 8:
Teach-In
Fig. 44: Status of Teach-In
Possible choices:
•
Missing: A Teach-In has not been executed (only for NF)
•
Without: There is no Teach-In existing, unit can be operated anyhow (only for FE)
•
OK: A Teach-In was already executed
•
On: Performs a Teach-In
Note
In order to enable measuring with a particular parameter set, a
Teach-In must have been executed.
If no Teach-In was executed or if a Teach-In has to be repeated, the menu option “on“ has to be
selected. Then the following two messages appears in succession:
Fig. 45: Removing sheet for Teach-In
Fig. 46: Attaching a sheet for Teach-In
Now a sheet with the nominal thickness must be presented to the sensor.
Attention
A sheet must be attached to the sensor without air gap. The position
of the sheet during Teach-In must be the same as for normal
operation. Otherwise the measurement result could be faulty.
Then press the ENTER key.
During the Teach-In procedure the following message is displayed:
Fig. 47: Display during the Teach-In
Note
Fault messages are described in chapter 9. “Fault messages ".
program par. 8
please wait
program par. 8
remove sheet!
program par. 8
attach sheet!
program par. 8
Teach-In: on