Maintenance
ROLAND ELECTRONIC GmbH · Otto-Maurer-Str. 17 · DE 75210 Keltern · Phone +49 (0)7236-9392-0 · Fax +49 (0)7236-9392-33
107
For a 2PW control unit with up two sensors:
1
Verify that no sheet is in front of the sensor resp. the sensors
2
Press the
MENU
key, switch to
“change“
and confirm with the
ENTER
key
3
Enter the password
4
Switch to
“program parameter“
and confirm with the
ENTER
key
5
Verify that no sheet is in front of the sensor resp. the sensors
6
Switch to
“zero adjust“
and confirm with the
ENTER
key
7
Select
“zero adjust: separate“
or
“zero adjust: common“
7.1
If “zero adjust: separate“, then confirm with
ENTER
key
7.1.1 Display: “zero adjust S1: on / remove sheet“
If zero adjusts desired, then
ENTER
key, or
if zero adjusts not desired, then activate “off“ and confirm with ENTER
7.1.2 Display: “zero adjust S2: off“
If zero adjusts not desired, then press
ENTER
key, or
if zero adjusts is desired, then activate “on“ and press ENTER
7.2
If “zero adjust: common“, then confirm with
ENTER
key, all connected sensors are
zero adjusted.
8
The display “zero adjust: active“ appears, and then “zero adjust: off“
9
Leave programming with the
MENU
key.