test ports. It is recommended that you preset the R&S
ZND to start from a well-defined
instrument state.
DUT
1
2
VNA
1
2
1. Connect the DUT between test ports 1 and 2 of the network analyzer as shown
above.
2. Use the [PRESET] key to restore a well-defined instrument state.
The analyzer is now set to its default state. The default measured quantity is the
transmission S-parameter S
21
.
Select TRACE – [TRACE
CONFIG] and use the control elements in the "Traces" soft-
tool tab if you wish to create additional traces and diagrams.
6.1.2
Selecting the sweep range and other parameters
After a system preset the display shows a diagram with a dB magnitude scale, and the
S-parameter S
21
is selected as a measured quantity. This S-parameter is the forward
transmission coefficient of the DUT. It is defined as the ratio of the transmitted wave at
the DUT's output port (port no. 2) to the incident wave at the DUT's input port (port no.
1).
Transmission S-parameter measurement