Instrument Tour
R&S
®
RTO2000
20
Getting Started 1332.9719.02 ─ 08
9 = Input channels
10 = [Aux Out] connector
11 = Connectors for probe compensation and grounding
12 = USB connectors
13 = [Power] key
Instrument damage caused by cleaning agents
Cleaning agents contain substances such as solvents (thinners, acetone, etc.), acids,
bases, or other substances. Solvents can damage the front panel labeling, plastic
parts, or screens, for example.
Never use cleaning agents to clean the outside of the instrument. Use a soft, dry, lint-
free dust cloth instead.
3.1.1
Input Connectors
The R&S
RTO has two or four channel inputs to connect the input signals using active
and passive probes.
The input connectors are provided with a special Rohde
&
Schwarz active probe inter-
face, and they are BNC compatible. Thus, the instrument can automatically detect pas-
sive probes with standard BNC connector and active Rohde
&
Schwarz probes having
the Rohde
&
Schwarz probe interface.
The input impedance is selectable, the values are 50
Ω
and 1
M
Ω
.
Risk of electrical shock or fire
Voltages higher than 30
V
RMS or 42
V
peak or 60
V
DC are regarded as hazardous
contact voltages. When working with hazardous contact voltages, use appropriate pro-
tective measures to preclude direct contact with the measurement setup:
●
Use only insulated voltage probes, test leads and adapters.
●
Do not touch voltages higher than 30
V
RMS or 42
V
peak or 60
V
DC.
Front Panel
Summary of Contents for R&S RTO2000 Series
Page 1: ...R S RTO2000 Oscilloscope Getting Started Getting Started Version 08 1332971902 P C2...
Page 6: ...Contents R S RTO2000 6 Getting Started 1332 9719 02 08...
Page 34: ...Instrument Tour R S RTO2000 34 Getting Started 1332 9719 02 08 Keys and Controls...
Page 72: ...Trying Out the Instrument R S RTO2000 72 Getting Started 1332 9719 02 08 Saving Data...