Setting Up the Instrument
R&S
®
RTO2000
101
Getting Started 1332.9719.02 ─ 08
Two connector pins are located on the front panel. The right pin is on ground level. The
left pin supplies a square wave signal with 1 kHz for low frequency probe compensa-
tion.
1. Connect the BNC connector of the probe to input [CH1].
2. Connect the probe's ground connector to the right compensation pin, and the tip
with the left pin.
3. Press [Autoset].
A square wave appears on the display.
4. Adjust the compensation trimmer of the probe to optimum square wave response.
For details, refer to the documentation of your probe.
Adjusting Passive Probes