Instrument Tour
R&S
®
RTO2000
23
Getting Started 1332.9719.02 ─ 08
USB
Two USB type A connectors that comply with standard USB 3.0. They are used to con-
nect devices like keyboard, mouse, printer, and flash drive to store and reload instru-
ment settings and measurement data.
Note:
Electromagnetic interference (EMI) can affect the measurement results. To avoid
any impact, do not use USB connecting cables exceeding 1
m.
LAN
8-pin RJ-45 connector used to connect the instrument to a Local Area Network (LAN).
It supports up to 1000 Mbit/s (10/100/1000BASE-T Ethernet).
USB Device
USB 3.0 interface of type B (device USB), to be used for remote control of the instru-
ment.
DisplayPort
DisplayPort connector for an external monitor or projector. It supports DisplayPort ver-
sion 1.1a.
DVI-D
Digital connector for an external monitor or projector. The monitor shows the complete
content of the instrument's screen.
[Ext Trigger In]
The BNC connector for external trigger input is used to control the measurement by an
external signal. The input impedance can be selected in the trigger configuration, the
values are 50
Ω
and 1
M
Ω
. The trigger level can be set from -5
V to 5
V. The maximum
input voltage is 30
V
RMS at 1
M
Ω
input impedance and 7
V
RMS at 50
Ω
input impe-
dance.
Trigger Out
The BNC connector for external trigger output is used to provide the internal trigger
signal of the oscilloscope to trigger other instruments for synchronized measurements.
When a trigger occurs, the R&S
RTO creates a pulse of 5
V with a source impedance
of 50
Ω
and delivers it to the external trigger output. The instrument can also send the
pulse on mask test violation or violation of measurement limits and margins.
If the connector is terminated with 50
Ω
, the signal level is 2.5
V (50
mA). With 1
M
Ω
termination, the level is 5
V. A short-circuit of the connector to ground creates current
of 100
mA.
To enable the trigger out signal, select "Trigger" menu > "Ctrl/Action". Here you also
adjust polarity, delay, and length of the pulse. The default is a positive pulse of 100
ns.
The minimum delay is 800
ns.
[RTO MSO]
Mixed Signal Option, input for digital signals (parallel buses). The hardware module
and logic probe come with option R&S
RTO-B1. The module provides connectors for
two logical probes with 8 digital channels each (D0 to D7 and D8 to D15).
Rear Panel
Summary of Contents for R&S RTO2000 Series
Page 1: ...R S RTO2000 Oscilloscope Getting Started Getting Started Version 08 1332971902 P C2...
Page 6: ...Contents R S RTO2000 6 Getting Started 1332 9719 02 08...
Page 34: ...Instrument Tour R S RTO2000 34 Getting Started 1332 9719 02 08 Keys and Controls...
Page 72: ...Trying Out the Instrument R S RTO2000 72 Getting Started 1332 9719 02 08 Saving Data...