Instrument Tour
R&S
®
RTO2000
21
Getting Started 1332.9719.02 ─ 08
Risk of injury and instrument damage
The instrument is not rated for any measurement category.
Make sure that the input voltage on
channel inputs
does not exceed 200
V
peak,
150
V
RMS at 1
M
Ω
input impedance and 5
V
RMS at 50
Ω
input impedance.
Transient overvoltages must not exceed 200
V
peak.
When performing measurements in circuits with transient overvoltages of category II,
III or IV circuits, make sure that no such overvoltages reach the R&S
RTO input.
Therefore, use only probes that comply with DIN EN 61010-031. When performing
measurements in category II, III or IV circuits, it is mandatory to insert a probe that
appropriately reduces the voltage so that no transient overvoltages higher than
200
V
peak are applied to the instrument. For detailed information, refer to the docu-
mentation and safety information of the probe manufacturer.
Explanation: According to section AA.2.4 of EN 61010-2-030, measuring circuits with-
out any measurement category are intended for measurements on circuits which are
not connected to the mains system.
3.1.2
Other Front Panel Connectors
Besides the input connectors, the instrument has USB connectors and probe compen-
sation connectors at the front panel.
[USB]
Two USB type A connectors that comply with standard USB 2.0. They are used to con-
nect devices like keyboard, mouse, printer and USB flash drive.
Note:
Electromagnetic interference (EMI) can affect the measurement results. To avoid
any impact, do not use USB connecting cables exceeding 1
m.
Probe Compensation
Probe compensation terminal to support adjustment of passive probes to the oscillo-
scope channel.
Protective earth conductor for grounding the instrument.
Square wave signal for probe compensation with 1
kHz and 1
V
pp
.
Ground connector for probes.
[Aux Out]
Output of the internal calibration signal, if the signal is configured to external destina-
tion.
3.2
Rear Panel
shows the rear panel of the R&S
RTO.
Rear Panel
Summary of Contents for R&S RTO2000 Series
Page 1: ...R S RTO2000 Oscilloscope Getting Started Getting Started Version 08 1332971902 P C2...
Page 6: ...Contents R S RTO2000 6 Getting Started 1332 9719 02 08...
Page 34: ...Instrument Tour R S RTO2000 34 Getting Started 1332 9719 02 08 Keys and Controls...
Page 72: ...Trying Out the Instrument R S RTO2000 72 Getting Started 1332 9719 02 08 Saving Data...