Adjust the Lag setting
In some situation (e.g. gastroparesis) it may be helpful to start a
bolus after you started eating. With the Lag setting you can
specify a delay between programming a bolus and the actual
beginning of the bolus delivery. Discuss using the Lag setting
with your healthcare team.
The
Therapy settings
menu appears.
Press
z
to move to
Bolus settings
.
Press
o
to select.
From the
Settings
menu press
z
to move to
Therapy settings
.
Press
o
to select.
The
Bolus settings
menu appears
Press
z
to move to
Lag setting
.
Press
o
to select.
The
Lag setting
screen appears.
Press
x
or
z
to move to
On
or
Off
as
needed.
Press
o
to select.