Table 202: Electrostatic discharge (ESD) and power frequency magnetic fields (all devices)
Guidance and manufacturer’s declaration – electromagnetic immunity
The Accu-Chek Insight insulin pump is intended for use in the electromagnetic environment specified below. The customer or the
user of Accu-Chek Insight insulin pump should assure that it is used in such an environment.
Immunity test standard
IEC 60601 test level*
Compliance level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast transient /
burst
IEC 61000-4-4
±2 kV for power
supply lines
Not applicable
Surge
IEC 61000-4-5
±1 kV line to line
±2 kV line to earth
Not applicable
Voltage dips, short
interruptions and voltage
variations on power
supply lines
IEC 61000-4-11
<5 % UT (0,5 cycle)
40 % UT (5 cycles)
70 % UT (25 cycles)
<5 % UT for 5s
Not applicable
Note:
UT is the A.C. mains voltage prior to application of the test level.
Summary of Contents for Accu-Check Insight
Page 3: ...Pump details and accessories Adapter tubing Battery Battery cover Cartridge Pump...
Page 4: ......
Page 8: ......
Page 62: ...52...
Page 76: ...66...
Page 96: ...86...
Page 140: ...130...
Page 146: ...136...
Page 159: ...149 Appendices...
Page 160: ...150...
Page 181: ...C 171...
Page 185: ...D 175 Icon Meaning Radio button selected Radio button not selected Icon on Icon off...
Page 205: ...195...
Page 211: ...201...