9.2 Damage to your pump
Dropping the pump could damage it and cause the waterproof
seal to crack. Use the specially designed pump carrying system
to help avoid dropping your pump. See the Accu-Chek sterile
products and accessories catalogues and brochures for more
information, or contact your local pump support (see inside back
cover).
If you drop your pump:
j
Check that all the connections in your infusion set are still
tight, and if they are not, reconnect or tighten them
j
Check your pump and its sterile products, and accessories for
chips and cracks
j
Change the cartridge if it is damaged
When you have completed these checks, take out and reinsert
the battery so that the self-test is started.
9.3 Repair of your pump
Do not perform any service or repair on your pump. Do not use
any lubricants on the pump mechanism. If you have questions,
contact your local pump support (see inside back cover).