3
45
Your pump cannot detect infusion set leakage. Inspect all parts of
your infusion set regularly during the day and before you go to
sleep. If there seems to be any loss of insulin and all parts are
correctly tightened, look for and immediately replace any leaking
component. Check your blood glucose level immediately, as
insulin delivery has been interrupted. Take appropriate action
according to your doctor’s or healthcare team’s instructions.
Always disconnect your infusion set before changing a cartridge
or filling the infusion set. Otherwise, you risk uncontrolled insulin
delivery into your body. Always follow the instructions for the
infusion set and cartridge you are using.
3.4 Changing the transfer set (adapter
and tubing)
Change the transfer set (adapter and tubing):
j
As directed by the instructions accompanying the infusion set
j
When the infusion site shows signs of irritation or infection
j
If the pump displays
Maintenance M24: Occlusion
j
If you think that there might be an occlusion in the tubing (even
without an alarm)
j
Every time you change the cartridge
Note
You can also review how to change the
cartridge. See chapter 2.3, page 26
Summary of Contents for Accu-Check Insight
Page 3: ...Pump details and accessories Adapter tubing Battery Battery cover Cartridge Pump...
Page 4: ......
Page 8: ......
Page 62: ...52...
Page 76: ...66...
Page 96: ...86...
Page 140: ...130...
Page 146: ...136...
Page 159: ...149 Appendices...
Page 160: ...150...
Page 181: ...C 171...
Page 185: ...D 175 Icon Meaning Radio button selected Radio button not selected Icon on Icon off...
Page 205: ...195...
Page 211: ...201...