Accuracy of flow rate for both
Accu-Chek Insight Insulin Pump
Cartridge and NovoRapid
®
PumpCart
®
Trumpet curve plotted from data after the end of the
stabilization period
The trumpet curve shows the accuracy of the delivery rate in
relation to the observation period.
8
0
–4
–6
–8
0
60
120
180
300
660
780
960
240
360
840
600
840
4
2
420
480
720
900
–2
6
–5.2000
–0.2000
0.3200
0.8909
2.3636
2.7200
1.7960
3.8000
5.6000
1.1579
2.0842
1.4774
1.6581
percentage Error of flow (%)
observation window duration (min)
minimum measured variation
maximum measured variation
overall percentage Error
* The measurements were made at a medium basal rate of 1.0 U/h according to IEC 60601-2-24:1998 with an Accu-Chek Insight Adapter and Tubing transfer set
100 cm at room temperature.
The maximum deviation of the delivered amount (overall mean
percentage flow Error) for U100 insulin is ≤ ±5 %*.