46
Turn the adapter with the tubing anti-
clockwise.
1
2
When you reach an angle of about 120°,
remove the adapter with the cartridge.
3
The
Rewind piston rod?
screen appears.
Press
o
to select
yes
.
Note
Disconnect and remove the infusion set’s head set and
insert a new one at a new site as described in the infusion
set’s instructions.
Change the transfer set (adapter and tubing)
w
WARNING
Inspect your infusion site at least twice each day for irritation
or infection. Signs of infection may include, but are not limited
to: pain, lumps, redness, heat, or discharge. If you see
redness or swelling, change the infusion set
’
s head set and
the infusion site immediately and contact your doctor or
healthcare team.
Change the head set according to the instructions for the
infusion set you are using and your doctor
’
s or healthcare
team’s recommendations.