A
165
Start-up graph over the stabilization period
The start-up graph shows changes in the flow rate over the
stabilization time.
flow rate (ml
/ h)
0.012
0.006
0.004
0.002
0.000
0
120
240
360
600
1080
1200
1440
480
720
840
960
1320
0.010
0.008
time (min)
real flow
target flow
* The measurements were made at a medium basal rate of 1.0 U/h according to IEC 60601-2-24:1998 with an Accu-Chek Insight Adapter and Tubing transfer set
100 cm at room temperature.
** The measurements were made according to IEC 60601-2-24:1998 with a Accu-Chek Insight Adapter and Tubing transfer set 100 cm at room temperature.
Influence of height on delivery accuracy
The maximum deviation of the delivered amount (overall mean
percentage flow Error) for U100 insulin is ≤ ±10
%* where your
Accu-Chek Insight insulin pump is located ≤ 0.8 m (2.62 feet)
above/below the infusion site.
Bolus
For U100 insulin the maximum deviation of a maximum bolus is
≤ ±5 % and the maximum deviation of a minimum bolus is
60 %.**