A
153
General Technical Data
Basal rate
Min. 0.02 U/h, max. 25 U/h.
Bolus
The max. bolus amount per delivery is 25 insulin units. The bolus amount for the Quick
Bolus is adjustable in increments of 0.1, 0.2, 0.5, 1.0 and 2.0 units. For the Standard
Bolus, Extended Bolus, and Multiwave Bolus, the amount is adjustable in increments of
0.05 units. The duration of the Extended Bolus and the Multiwave Bolus is adjustable in
intervals of 15 minutes up to 24 hours.
Temporary Basal Rate
Adjustable in 10 % increments, 0 – 90 % for decreases, 110 – 250 % for increases. The
duration is adjustable in 15 minute intervals, up to a maximum of 24 hours. The last
programmed duration is given by default for the next Temporary Basal Rate change
programmed.
Bolus delivery speed
very slow = 3 U/min, slow = 6 U/min, moderate = 9 U/min, standard = 12 U/min
Priming speed
12 U/min