Configuration
Calibrate current interface when
value is selected
Enter the value that corresponds to 20 mA.
Press the [OK] key.
Calibrate current interface
Enter the value that corresponds to 20 mA.
Here you can enter the measured value to which the current
interface is to be scaled.
We recommend setting the value for 20 mA to the end of the
reagent measuring range, but values below this can also be
selected. For more information, see page 38.
You will see an input keyboard on the right. Now use the
arrow keys [
◀
], [
▶
], [
▲
] and [
▼
] to move the black cursor to
the desired digit and press the [OK] key.
The selected digit appears in a frame on the left.
Repeat the entry until the desired number appears in the
frame. Now move the cursor to the OK field on the keyboard
and press the [OK] key.
You will see the following information on the left:
Current:
currently programmed value
Minimum:
smallest adjustable value
Maximum:
largest adjustable value
If the value at "Current" corresponds to your wish, you do not
need to enter the number again and can immediately move
the cursor to the OK field in the keyboard and confirm with
the [OK] key.
Wizard
The configuration is complete. The wizard is closed.
Press the [OK] key.
This completes the configuration of the unit.
If necessary, individual settings can also be made without the wizard.
Programming: Menu > Settings