Switch for external analysis start
In addition to the possibilities of starting an analysis time-dependently, it is possible to trigger
additional analyses via an external button. This can be a potential-free button or the relay output of
a PLC control or softening control.
The use of a switch at the input is intended as an addition to the normal time interval. The analyser
works in a set time interval, via the switch a signal can be delivered to the input and thus an
additional analysis can be started.
(When using this function, the Sycon must be in automatic mode).
An additional analysis can also be triggered by pressing the [OK] key for 3
seconds.
With a permanently connected input contact in analysis start mode, analyses
are performed permanently.
Programming: Menu > Settings > Inputs > Input 1 > Start analysis
Figure 17Three possible connection variants around the analyser (1)
Connection of external components