
July 2022
© RLS Wacon analytics GmbH +49 (0)5121 281260
Pag
e
56
Subject to
change without notice
Configuration
If flow monitor is selected
Is the flow monitor a NC or NO type?
[NO] Analysis interval paused: Next analysis start is waiting
for a closed input contact.
[NC] Analysis interval paused: Next analysis start is waiting
for an open input contact.
Make your selection and confirm with the [OK] key.
(When using this function, the Sycon must be in automatic
mode).
If interval reset is selected
[NO] Analysis interval active: Analysis interval is stopped by a
closed input contact.
[NC] Analysis interval active: Analysis interval is stopped by
an open input contact.
Make your selection and confirm with the [OK] key.
If the analysis interval is stopped, then the analyses are
stopped immediately and paused as long as the input contact
remains closed/open.
As soon as the input contact is opened/closed, the analysis
interval begins. However, since there is no display value in the
display yet, the first analysis is carried out after 1 min before
the set analysis interval (e.g. 5 min) begins.
(When using this function, the Sycon must be in automatic
mode).